9.1 Items to be Prepared by User
9 - 2
9.1 Items to be Prepared by User
For carrying out instrument diagnosis, you must prepare the
following items beforehand.
(1) Single-element mercury (Hg) hollow cathode lamp (part No.
207-2007)
(2) Single-element copper (Cu) hollow cathode lamp (part No.
208-2011)
(3) Copper (Cu) standard solution
(4) Graphite cuvette (Pyro Tube HR)
NOTES
: 1. Any hollow cathode lamp other than made by
Hitachi High-Technologies cannot be used for
instrument diagnosis. Use of other maker’s
hollow cathode lamp may result in failure to
achieve exact diagnosis or a shortening of lamp
life.
2. If a hollow cathode lamp whose service life has
already been expired is used, instrument
diagnosis may not be carried out correctly.
To avoid this, be sure to use a hollow cathode
lamp that has not yet reached the end of its
service life.
3. Use a new standard solution prepared at the time
of inspection.