5.1.2
5 - 7
Fig. 5-5 Working-curve Table Window
(8) In the [STD addition] (Fig. 5-5 (1)) field, select [Inner] and
make entry for [Blank Cup No.].
Use of the default value (No. 59) does not constitute any
problem.
“Blank” hereby indicates a blank sample for determining the
extrapolation value of working curve. Prepare what
resembles humoral of unknown samples, for example, the
diluents of the unknown samples. Or use purified water if
the former is unavailable.
(9) Verify the conditions, place standard samples, unknown
samples and blank samples in the sample rack, and then
start measurement.