3.1.9
3 - 23
Click the [Report Format] button (<9> in Fig. 3-1) to display the
Report Format dialog. Auto-printing at the end of measurement
and the items to be printed can be selected here.
When auto-printing is not specified, printing can be made upon
selecting the desired items at the end of measurement.
Given here is a simple example where auto printing is not made.
In this case, the procedure is advanced without setting anything
on the Report Format dialog.
For the method of report setting, refer to “4.4 Saving and
Printout of Result.”
Automatic saving of analysis results
After completion of the measurement sequence, the
measured data are saved. Enter the name of the folder to
save to and the name of the file in the Filename field. The file
name contains a date and time number, preventing the
previous data from being overwrtitten.
Automatic reporting
If this box is checked, the measurement results will be printed
in the specified format after completion of the measurement
sequence. Formats include print (paper), text and Microsoft
Excel.
When auto-printing is not specified, printing can be made upon
selecting the desired items at the end of measurement.
Given here is a simple example where auto printing is not made.
In this case, the procedure is advanced without setting anything
on the Report Format dialog.
Click the [Parameter List] button (<10> in Fig. 3-1), and a list of
conditions will be displayed to allow check of the main set
parameters. Display is changeable by the [Conditions for Each
Element] and [Common Conditions] tabs at top of this window.
In particular, confirm that the number of measurement
repetitions and other parameters required to be set for each
element are set properly (this window is used only for review and
does not allow any editing).
3.1.9 Setting
Report
Format
3.1.10 Check of
Analytical
Conditions
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...