10.1 Setting of Temperature Program Test Conditions
10 - 6
Fig. 10-5 Dry Test Window
Start
Temp.
End
Temp.
First measurement
80 °C
150 °C
Second measurement
110 °C
150 °C
Third measurement
130 °C
150 °C
[Dry Test] is executed up to the [Start Temp.]
given by “[End Temp.] - 20°C”.
Therefore, the third program starts not at 140°C
given by adding the final increment to 110°C but at
130°C determined by “[End Temp.] - 20°C”.
Ramp Time (s)
Enter a heating time when the starting temperature
reaches the ending temperature in the drying stage.
Usually enter a value of about 40 s (second).
Input range: 10 to 100 s
Gas Flow (mL/min)
Enter a carrier gas flow rate in the drying stage.
Usually there is no need for changing 200 mL/min.
Input value: 0, 10, 30, 200 mL/min
Gas Type
Select the desired kind of carrier gas in the drying stage
(between Normal and Alternate).
Usually select Normal (Argon Gas).
NOTE
: For using the Alternate gas, a stage for replacement
with the normal gas is necessary.
Set the [Dry addition] or [Ash addition].
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...