1.3 Supply of Gases
1 - 4
1.3 Supply of Gases
(1) Prepare the fuel and auxiliary gases in accordance with the
atomizer to be used.
Atomizer
Combination of Fuel and Auxiliary Gases
Standard burner
Air-acetylene, air
High-temperature
burner
Nitrous oxide-acetylene, air-acetylene, air
Hg cell
Air
The high-temperature burner and Hg cell are options.
Refer to their respective instruction manuals for using
these items.
When air alone is employed for the standard or high-
temperature burner, it is used for cleaning or drying of
the burner head.
(2) When a compressor is used for air, turn on power to the
compressor. For compressed gases, open the main valve
of each gas cylinder.
NOTE
: Solvent such as acetone is included in the acetylene
cylinder. To prevent deterioration of gas piping or
damage to the instrument due to the outflow of solvent,
replace the cylinder with a new one when gas pressure
(primary pressure) degreases to 0.5 MPa. Firmly
fasten the main cock of each cylinder after use.
(3) Set the secondary pressure (supply pressure) of each gas
as follows.
Gas Type
Secondary (supply) Pressure
Acetylene
0.09 MPa (0.9 kgf/cm
2
, 13 psi)
Nitrous oxide
0.4 MPa (4 kgf/cm
2
, 60 psi)
Argon
0.5 MPa (5 kgf/cm
2
, 70 psi)
Air
0.5 MPa (5 kgf/cm
2
, 70 psi)
(4) Open the supply valve of each gas to the instrument.
NOTE
: Open the air valve after the pressure of the
compressor reached 0.2 MPa, otherwise the air will be
released from an air filter installed beside the
spectrophotometer.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...