5.3
5 - 11
5.3 Measurement under the Same Conditions (Copy Method)
A new measurement can be carried out under the same
conditions as those of the presently open analytical result by
utilizing the Copy Method function.
(1) For using the working curve that has been drawn in the
previous measurement, open the Data Process window,
and click the [Copy Method] button on the toolbar.
Fig. 5-9 [Copy Method] Button
In case the instrument is now in the stop status due to
clicking of the [Stop] or [Pause] button during measurement,
you should click the [Close Sequence] button. Then, the
Data Process window will appear.
(2) In response to the confirmation message, click the [Yes]
button.
(3) Display changes to the Method window.
Change conditions such as the number of unknown
samples if necessary. Then, verify the method and start
measurement.
NOTES
: 1.
Lamp numbers are also copied. Check the
installation number of a hollow cathode lamp again.
2. Burner head heights are also copied. Adjust the
front height of the frame part with an adjusting
screw as required.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...