5 - 1
5 MAINTENANCE
Continuous use of the instrument without periodic inspection and
maintenance may result in degradation of components that can
cause some serious problems, including gas leakage, liquid leakage,
electricity leakage, or ignition.
Perform inspection and maintenance according to the specified
intervals regardless of the frequency of use.
5.1 Replacement of Pump Tube
When a pump tube is deteriorated or damaged (torn), replace with a
spare. The procedure is instructed below.
(1) Release the support clamp lever by raising it.
(2) Turn the tube support down.
(3) Disconnect the pump tube from the rotor.
Fig. 5-1 Removal of Pump Tube
(4) Disconnect the old tube from the nipple.
(5) Connect a new tube to the nipple using the furnished sandpaper.
Fig. 5-2 Connection of Pump Tube and Nipple
Sandpaper
Pump tube
Nipple
Support clamp
lever
Nipple
Pump tube
Tube support
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...