3.1 Generation of Measurement Conditions
3 - 10
Measurement Time
Obtain the absorbance data for the time set from the
start of measurement and calculate the level of
absorbance. This is normally set at about 5 seconds.
Input range is from 0.5 to 20 seconds.
Auto-Start
This is a function for automatically starting measurement
upon detecting the rise of the atomic absorption signal at
sample introduction. This function does not work for
samples such as a blank solution of which absorbance is
almost no different from the baseline. In that case,
manually click [Start].
Continuous Measurement
When this check box is turned on, measurement of
samples will be performed through continuous injection
of sample and data will be acquired continuously
according to the number of repetitions, without rinsing
with pure water in between measurements. For using
this method, click the check box to put a check mark
there.
(Example) Case of three repetitions
Continuous introduction ON
Continuous introduction OFF
Blank
1st
2nd
3rd
Blank
Blank
1st
2nd
3rd
Blank
Blank
Blank
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...