5.8 Emission Measurement
5 - 20
Element
Wavelength
(nm)
Description
B 518.0
Nitrous oxide - acetylene flame used (high-
temperature burner)
Ba
553.6
Air - acetylene flame used (standard burner)
Ca
422.7
Air - acetylene flame used (standard burner)
In
451.1
Air - acetylene flame used (standard burner)
K
766.5
Air - acetylene flame used (standard burner)
Li
670.8
Air - acetylene flame used (standard burner)
Mn
403.1
Air - acetylene flame used (standard burner)
Na
589.0
Air - acetylene flame used (standard burner)
Sr
460.7
Air - acetylene flame used (standard burner)
(3) In the Instrument Setup window, specify the following
items.
Wavelength
Wavelength Setting
Time Constant
(The time constant should be one second or more.)
PMT Voltage
Absorbance Decimal Place
Calculation Mode
Slit Width
Fig. 5-17 Instrument Setup Window