1.3 Supply of Gases
1 - 4
1.3 Supply of Gases
(1) Prepare the gas which matches the heating program to be
used.
Gas Type
Gas
Normal Argon
Alternate
Mixture of argon + oxygen (5 to 10%)
Alternate gas is usable only when the optional alternate
gas unit is provided.
NOTES
: 1. When the primary pressure of argon gas cylinder
falls close to 0.7 MPa (7 kgf/cm
2
, 100 psi), the
cylinder should be replaced with a new one. If
gas falls short of an adequate level during the
course of analytical operation, an error is detected
and measurement is interrupted.
2. When ordering argon gas, specify “dry argon gas
(purity 99.95% or higher)”. Argon gas having a
lower purity may contain moisture.
The moisture contained in argon gas will be
heated at a high temperature and react with the
graphite cuvette, so the service life of the cuvette
may shorten.
(2) Open the main valve of each gas cylinder.
(3) Set the secondary pressure (supply pressure) of each gas
as follows.
Gas Type
Secondary (Supply) Pressure
Argon
0.5 MPa (5 kgf/cm
2
, 70 psi)
Alternate
0.5 MPa (5 kgf/cm
2
, 70 psi)
(4) Open the supply valve of each gas to each gas instrument.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...