1 - 2
1.2 Required Equipment
1.2.1 Power Supply
The specifications of the power supply for this system are given
in Table 1-1.
Table 1-1 Power Supply Specifications
Part No.
Voltage
Allowable
Voltage Fluctuation
Capacity Frequency
7J1-8821 115
V
±10 %
50 VA
50/60 Hz
(common)
7J1-8822 220
V
7J1-8823 230
V
7J1-8824 240
V
7J1-8827 220
V
7J1-8828 220
V
Grounding
:
Connect the instrument to a ground terminal with a
grounding resistance of 100 ohms or less.
1.2.2 Gases
Prepare compressed air (compressor), acetylene and argon gases.
The specification of the Argon gas for this system are given
in Table 1-2.
Table 1-2 Argon gas
Specification
Gas
Pressure Regulator
Supply Pressure
(secondary pressure setting)
Argon
Primary : 25 MPa
Secondary: 1 MPa
Full scale
500 kPa
Attach a stop valve to the terminal of the pipe in the measuring room.
The details refer to the ZA3000 series Maintenance Manual.
NOTE:
Operation with an improper gas pressure may result in a
performance degradation of the instrument.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...