10.1 Setting of Temperature Program Test Conditions
10 - 2
10.1 Setting of Temperature Program Test Conditions
For using the function for supporting automatic generation of
temperature program, setting is required on the [Method] -
[Analytical Condition] window. When this function is used, usual
quantitative analysis will not be carried out.
Unknown samples should be measured after the temperature
program obtained by this function has been reflected on
analytical conditions.
(1) After entering necessary conditions in the [Elements] and
[Instrument Setup] windows, and set the UNK replicate to
twice or more in the [Working curve table] and enter the
unknown sample ID to be evaluated in the [Sample table]
and click the [Temp. Program Development] check box in
the [Analytical Condition] window to put a check mark. In
this function, multiple elements are measurable.
NOTE
: When checking the [Temp. Program Development], all
the elements set for [Elements] will become the objects
of the automatic temperature program generation.
For using this supporting function, select only the
desired elements.
Fig. 10-1 Analytical Condition Window
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...