4 - 4
4.4 Setting of Reagents
Place the following reagent bottles on the reagent tray of the HFS-4
and insert each solution tube into the bottle according to the
identification marker.
(1) 1% sodium borohydride solution
Green identification marker.
(2) 1.0 mol/L hydrochloric acid
Red identification marker.
(3) Reducing
agent
solution
Blue identification marker.
(4) Sample
Without identification marker. Another pure water bottle should
be prepared for washing the suction tube.
4.5 Measurement Using Four Liquids
4.5.1 Preparation for Measurement Using Four Liquids
(1) As preparation for the measurement, set the conditions of the
AAS main unit. Refer to “3.4 Setting of Conditions”.
(2) Set the absorption cell. For details, refer to “3.5 Setting of
Absorption Cell.”
(3) Set up the HFS-4 for the measurement.
Check that each of the four pump tubes is set in the roller
groove of the pump. Then, push the tube support onto the pump
tube and set the support clamp lever in position by lowering it.
NOTE 1:
Incorrect setting of reagent bottles or pump tubes can
lead to inaccurate measurement results.
NOTE 2:
Lower the support clamp lever securely so that it
becomes horizontal.
CAUTION
Your hand or hair may be sucked into the pump.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...