9
9 - 1
9. INSTRUMENT DIAGNOSIS
Instrument diagnosis function checks the performance of
instrument through measurement on the following items.
Wavelength accuracy
Baseline stability
Sensitivity
Repeatability
The performance of instrument is warranted to match the
product specifications through inspection at the factory. As the
instrument is used longer, however, some performance items
may be degraded due to deterioration of parts having a limited
service life (consumables) or for similar reasons.
Therefore, the performance needs to be checked.
Additionally, certain parts require maintenance with periodic
replacement or cleaning.
The method adopted in the instrument diagnosis function is a
modified version of the inspection method implemented at the
factory before shipment so that the customer can utilize it easily
to compare initial and current performance.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...