9.3 Operation
9 - 4
9.3 Operation
Follow the instructions below, assuming the instrument is
diagnosed on all items.
The lamps, standard solutions or graphite cuvette may not be
required if all diagnostic tests are not performed.
(1) Install the lamps into the lamp turret.
The mercury and copper lamps should be set in Nos. 8 and
7, respectively.
(2) Set the graphite cuvette (Pyro Tube C).
(1) Click the [Diagnosis] button in the toolbox to display the
Diagnosis window.
Fig. 9-1 [Diagnosis] Button
(2) Click the [Measurement Mode] button on the toolbar, or
[Measurement Mode] in the [Method Setup] menu.
The Measurement Mode dialog box opens.
In Measurement Mode, select [Graphite Furnace].
Fig. 9-2 [Measurement Mode] Button
9.3.1 Preparation
9.3.2 Setting
of
Conditions