6. USING THE QC FUNCTIONS
6 - 2
NOTES
: 1. For using the QC function, the following conditions
need to be set.
Use Saved
Working Curve
Determination ........... Unused
Mode ........................ Working Curve or
Simple STD Add.
Reslope .................... Unspecified
Conc. Times ............. 1
2. For the data measured by the QC function, the
following data reprocessing operations cannot be
specified. Otherwise, the data may change
causing a disagreement with the message in QC
measurement.
Change of working curve order, change of STD
concentration, deletion of sample, restore of
sample, exchange of samples, calculation mode,
sample blank, zero point shift of working curve,
baseline shift. However, if only Check Working
Curve is desired, deletion/restore/exchange of
unknown samples and sample blank are allowed.
3. For using the QC function, the Dil. Factor of
unknown sample should be set to 1 in the
Autosampler window.
4. When Simple STD Add. is selected in
Determination Mode, Check STD and Check
Recovery cannot be specified.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...