1.5.3 Setting of Four-liquid Flow Channel
Reconfiguration of the flow channel is required for measurement
using the four-liquid flow channel, in which a preliminary reducing
agent is automatically added by the pump.
Follow the procedures below to reconfigure the flow channel.
1.5.3.1 Setting of Reaction Coil
(1) Remove the screws from the right and left sides and the back of
the instrument and detach the cover.
Fig. 1-6 Internal Flow Channel
(2) Remove the tube with band marking #4.
Fig. 1-7 Internal Flow Channel after Removing Tube with Band Marking #4
Tube with band
marking #4