9. TROUBLESHOOTING
9 - 2
(cont’d)
Symptoms
Possible Causes
Check and Remedies
Rate of recovery
improper
Interference due to a
coexisting substance in
sample
Decrease the concentration of the coexisting
substance by means of dilution or extraction.
For some elements, add a matrix modifier.
Sample viscosity too high
Aspirate the sample, and check the degree
of aspiration. Decrease the viscosity of the
sample by means of dilution or
decomposition.
Baseline not stable
Lamp deteriorated
Check the displayed lamp spectrum, and
replace the lamp with a new one.
Lamp current improper
Adjust the lamp current properly.
Wrong lamp (element)
Check the lamp house, and replace the lamp
with a proper one.