6.4 Using the QC Sample (Check QC Sample)
6 - 10
6.4 Using the QC Sample (Check QC Sample)
A QC sample may be measured at a specified interval. If the
measured concentration is not within the permissible range,
remeasurement of working curve and unknown sample can be
performed as a failure action according to specification.
Other permitted actions are continuing (ignore failure) or ending
the sequence.
The permissible range is judged from the following expression.
Two different QC samples are supported, you can specify QC
Sample 1 and QC Sample 2.
(1) Click the check box for Check QC Sample.
Fig. 6-5 STD Check Window
(Measured concentration) – (QC sample concentration)
(QC sample concentration)
(Tolerance %)
100
6.4.1 Setting
Procedure