5.1.4
5 - 15
Fig. 5-13 Working-curve Table Window
(8) In the [STD addition] field, select [Inner] and make entry for
[Blank Cup No.].
Use of the default value (No. 59) does not cause any
problem.
“Blank” hereby indicates a blank sample for determining the
extrapolation value of working curve. Use the unknown
sample diluent or purified water if the former is unavailable.
(9) Verify the method, place standard samples, unknown
samples and blank sample in the sample tray, and then
start measurement.
NOTE
: For the first unknown sample, select the one that
contains a matrix constituent (co-existing substances)
having the average composition of all unknown
samples and whose absorbance is not extremely high.
(1)