EN
Solight 1T06 Breath Analyser User Manual
Thank you for buying our digital breath analyser, which measures residual alcohol content in your breath and indicates
the possible level of intoxication. The 1T06 model ranks among the professional measuring devices in its branch. It is
equipped with a measuring sensor relying on the Fuel Cell technology, which is based on an electrochemical reaction
of platinum and ethanol and is able to determine the alcohol content in breath with high accuracy of three decimal
places of PPT.
Instructions for use
1. Insert three 1.5V AAA size
alkaline
batteries into your device.
2. Prior to using the breath analyser, wash out your mouth with clean water. Do not smoke or drink alcohol for at
least 10 minutes prior to measurement.
!!! Measuring immediately or shortly after drinking alcohol is inaccurate and usually also causes damage to the
sensitive measuring sensor. Claim due to this cause would be unjustified.
3. Extend the mouthpiece using the slider at the top of the device. Only this expandable mouthpiece can be used
for testing. When testing several persons, for hygiene, put a blow-in tube onto the mouthpiece. Use a new blow-
in tube for each person.
4. Push power ON button to turn the device on. ALCO text will appear on the display.
5. By pushing the same button once again, you will start device heat-up. This stage includes an automatic testing
chamber cleaning, sensor activation and the device will be ready for use after approx. 30 seconds. “PH” text will
appear on the display, followed by a countdown sequence to zero - SH99 to 00 will appear on the display. At
the end of the countdown, an audible signal will sound and “READY” and “BLOW” texts will start to flash. A
countdown interval for blowing will appear on the display. If you do not start within approx. 30 seconds, the test
will be cancelled.
6. Insert the mouthpiece into your mouth and blow-in uninterruptedly for around 5 to 7 seconds.
7. Stop blowing after an audible signal. The device will now start analysing, which is indicated by “ANLY” text
flashing on the display.
8. The measured value will then appear on the display for 10 seconds and then gets stored into device memory.
Values are displayed with three decimal spaces; measurement deviation is ±0.008 ‰
A pre-set value of 0.2 ‰ will alert excess of the tolerated value limit for driving motor vehicles. “HI” and “Do
not drive” icon will start flashing on the display.
9. If the air flow is too short or too weak, “FLO” will appear on the display and the test must be repeated.
10. For new measurement, press button again.
The device stores 20 measured values in its memory.
11. The measured values can be displayed by pushing the Up/Down buttons on the side of the device.
12. After five subsequent measurements, leave the device at standstill for 5 minutes.
13. By pushing the button for 2 seconds you will turn the device off. The device will power off automatically after
approx. 30 seconds.
Blow-in
tube
Up/Down
buttons
Power
supply
On/Off
Extensible mouthpiece control
Blow
Do not
drive
Measure
ment
units
Wait
Ready
Alcohol content exceeds
pre-set limit
Battery
level