SAFETY - 6
SAFETY SUMMARY
List of WARNING Indications
Beware of High Voltage!
There is a danger of electric shock due to high voltage of
about 400 V. Touching the socket of the hollow cathode
lamp could result in fatal or serious injury.
Before replacing the lamp, be sure to turn OFF the lamp
current (click the lamp OFF button).
(Section 1 in this manual)
Beware of Toxic Gas!
Inhalation of toxic gas could be injurious to human health.
Inhaling the vapor emitted from the sample may cause
inflammation of the respiratory tract.
Be sure to operate the ventilating fan to discharge the vapor
emitted from the sample through the exhaust duct.
(Sections 1, 3 and 5 in this manual)
Explosion due to Backfiring!
If water is not supplied to the drain trap, there is a danger of
explosion due to backfiring, or ignition due to gas leakage.
Be sure to supply the drain trap with water.
(Sections 1 and 5 in this manual)
Explosion due to Acetylide Formation!
If samples containing a large amount of copper, silver or
mercury are atomized, it may generate metal acetylide and
cause explosion due to backfiring with louse noise. Do not
measure samples containing a large amount of copper,
silver or mercury.
(Quoted from JIS K 0120 General rules for
atomic absorption spectrometry)
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...