3 - 13
3.9 Measurement with Autosampler
(1) Exchange the Teflon tube of the sample suction tube for the
furnished autosampler tube (part No. 171-6335).
(2) One sample measurement will require a sample volume of about
10 mL. Prepare adequate volume of sample.
(3) Delay Time on the Analytical Condition window should be set to
about 80 s. Set an optimum delay time according to the
measurement of a sample solution.
NOTE 1:
When Se is measured after As measurement, the
absorbance of Se may decrease because of KI remained
in the flow channel. In case of measurement for both of
As and Se, measure Se first. When Se has to be
measured after As measurement, wash the flow channel
by aspirating HCl solution (e.g. 1.0 mol/L) for about 10
min.
NOTE 2:
When the measurement is continued for long time (more
than 1 h), the absorbance may decrease because of
absorption of analyte into dew inside of the silicone tube
(This will be prominent for Se). In this case, remove the
silicone tube and dry it, then attach again.