10.1
10 - 15
4. In a measurement immediately after a cuvette is
mounted, the reference signal may not stabilize
due to contamination of the cuvette.
After mounting a cuvette, carry out heating at the
highest temperature a few times.
The Temp. Program Development function should
be executed after stabilization of the reference
signal.
Detection Start
Enter the REF absorbance for detection of background
absorption.
Usually, the default value should be used.
Input range: 0.0001 to 1.0000 Abs
Detection End
Enter the REF absorbance at which it is judged that
background absorption ended after it had been detected.
Usually, the default value should be used.
Input range: 0.0001 to 1.0000 Abs
Ash addition
This function is provided for adding one more ashing stage
under the determined conditions after an ashing test and
then executing the temperature program.
When you apply check mark to the “Ash addition” check
box by clicking it, stage 6 (ashing) on the heating
temperature program will be implemented under the input
conditions.
Fig. 10-11 Ashing Test Window
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...