5.12 Measurement under the Same Conditions (Copy Method)
A new measurement can be carried out under the same
conditions as those of the presently open analytical result by
utilizing the Copy Method function.
(1) For using the working curve that has been drawn in the
previous measurement, open the Data Process window,
and click the [Copy Method] button on the toolbar.
Fig. 5-34 Copy Method Button
In case the instrument is now in the stop status due to
clicking of the [Stop] or [Pause] button during measurement,
you should click the [Close Sequence] button. Then, the
Data Process window will appear.
(2) In response to the confirmation message, click the [Yes]
button.
(3) Display changes to the Method window.
Change conditions such as the number of unknown
samples if necessary. Then, verify the method and start
measurement.
NOTE
: The lamp No. is also copied. Also re-confirm the
installation No. of the hollow cathode lamp.