5.4
5 - 25
NOTES
: 1. Samples may be contaminated as only the outer
wall of the nozzle is cleaned before aspirating the
next sample. Please be careful in using the
sequential injection. The sequential injection is not
recommended for measurement in the order of
multiple
μ
g/L(ppb), measurement of elements
susceptible to contamination such as Na and Zn,
etc., measurement of samples with highly
concentrated elements and measurement of
anything with large contamination factors.
2. The cleaning of the nozzle’s outer wall is not
designed to completely remove contamination.