3.1.3
3 - 3
(2) Click the elements to be analyzed in order on the Periodic
Table (<1> in Fig. 3-2). The selected elements will
change to green and will be added to the [Element-Order]
table (<2> in Fig. 3-2).
A maximum of 8 lamps can be set in the instrument, and
when using multi-element lamps, up to 12 elements can be
set. For the lamp number, input the number given in the
socket of the lamp installed in the main unit.
(3) To delete an analyte element, use the following steps.
Click an element in the periodic table.
Click the element highlighted in yellow-green to change
the color to white.
To uncheck [Measurement] in the [Element-Order], click
the check mark
in the [Measurement] column.
The row will remain in the table, but measurement will
not be carried out.
Click
[Delete].
Click [No.] in the [Element-Order] column for the
element to delete to change the color of the row of the
element to blue (<3> in Fig. 3-2). Click the [Delete]
button.
Click the [Delete All] button to delete all elements in the
[Element-Order] column.
(4) Click a [Lamp No.] in the [Element-Order] table and enter
the turret number of the lamps set for each element.
(5) Measurement is made in the order indicated at [No.].
To change the order, click the [No.] of the element to be
changed and when that row turns blue, shift it with [Up] or
[Down] button.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...