3.8.1
3 - 37
CAUTION
Beware of Intense Light!
An intense ultraviolet light is radiated from the flame in
flame analysis, which could damage the eyes.
Be sure to wear protective goggles when using the
instrument.
And do not detach the window at the front of the flame
section when using the instrument.
CAUTION
Strong Magnetic Fields!
The atomizer unit utilizes a magnet.
The strong magnetic field produced may cause damage to
a watch, magnetic card and so on.
Avoid bringing such objects near the magnet.
CAUTION
Deterioration of Burner Chamber by Chemicals!
A strong acid other than hydrochloric acid or hydrofluoric
acid may damage the burner chamber.
Particularly avoid using concentrated nitric acid,
concentrated sulfuric acid or aqua regia since they will
corrode the chamber.
For the acceptable solvents, see section 2 in the Maintenance
Manual.
(1) Inject the pure water, and click the [Ready] button in the
toolbox. The process will start from condition setting if the
hollow cathode lamp is not lighted.
Fig. 3-31 Ready Button
(2) Wait for five minutes while continuing pure water injection
for cleansing of the chamber.
3.8.1 Preparation
for
Measurement
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...