6.2 Using the Working Curve Range and RSD (Check Sample)
6 - 6
NOTES
: 1. When RSD of remeasurement was outside the
permissible range again, the next sample
measurement will be carried out.
2. When the determination mode is set to Simple
STD Add., the first unknown sample will be
excluded from the RSD limit check.
The following messages are printed in the report:
The measurement concentration exceeds the effective range
of the working curve. Dilute the sample and remeasure it.
The sample concentration is too high (it will be outside the
range of working curve even if it is diluted).
RSD exceeded the permissible range (go to the next step if
the remeasured value still exceeds the range).
6.2.2 Contents to be
printed on the
report