8.3 Flowing of Cooling Water
Supply cooling water to the instrument.
NOTES
: 1. A water flow of about 2 L/min is required within a
water pressure range from 35 to 50 kPa (about
0.35 to 0.5 kgf/cm
2
). If the temperature of
cooling water is much lower than the room
temperature, condensation of water vapor in the
air may occur on the quartz window of atomizer.
Water temperature should be limited within ±5 °C
of the room temperature.
2. Cooling water should be supplied shortly before
measurement for preventing condensation.