5.1.2
5 - 5
Fig. 5-4 Standards Table Window
(3) In [Determination Mode], select [STD Add.], and in [Order],
select [Linear].
(4) In the [Working Curve] column, make entries for [No. of
STDs], [STD Replicate], [STD Decimal Place], [STD Unit]
and STD [Conc.].
In the standard addition method, concentration is calculated
by extrapolation of a linear working curve. Therefore, STD
concentration should be set within a range where working
curve is neither concave nor convex. In addition, STD1
concentration is required to be 0.
(5) Make entries for [UNK Decimal Place] and [UNK Unit].
[UNK Replicate] in the standard addition method is fixed to
1 (once). Entry of another value will be neglected.
(6) Open the Sample Table window, and specify “1” as the
number of unknown samples and enter a sample name.
In the standard addition method, even if two or more
unknown samples are set up, just the first unknown sample
is measured.
(7) After confirming the conditions specified in the above steps,
start measurement.