5.1 Various Measurement Methods
5 - 12
This method has two steps. The standard addition method is
used for the first samples, and then the second samples onward
are measured by using (horizontally shifting the working curve)
the gradient of the working curve affected by the co-existing
substances. (See Fig. 5-10.) So, this method is equivalent to the
working curve generation with matrix-matched samples (matrix
matching method).
Fig. 5-10 Working Curve in Simple Standard
Addition Method
This method is effective for measuring a number of samples
similar in matrix composition.
The in-furnace addition method is a method of preparing and
measuring mixed samples (not mixed equally) in cuvettes by
preparing standard solutions and unknown samples separately
and adding each sample individually.
If mixing of unknown and standard samples in the furnace is
inadequate due to a reason such as high sample viscosity, the
ex-furnace addition method should be used.
(1) Starting from the Method window, make entries in the
[Measurement Mode] window, [Elements] window,
[Instrument Setup] window and [Analytical Condition]
window. For these entries, refer to Section 3.
(2) Click the [Standards Table] button to display the Standards
Table window (Fig. 5-11).
5.1.4 Simple
Standard
Addition Method
(in-furnace
addition)
A
x+2
A
x+1
A
x
-C
x
0
C
1
C
2
Concentration
C1:
Standard addition temperature for unknown
sample 1 (1)
C2: Standard addition temperature for unknown
sample 1 (2)
Ax: Absorbance of unknown sample1
Ax+1: Absorbance of unknown C1
Ax+2: Absorbance of unknown C2
x:
Reciprocal number of unknown sample1’s
concentration
Absorbance
Working curve for simple
standard addition method
(Unknown sample 1 and
onward)
Working curve for simple
standard addition method
(Unknown sample 2 and
onward)
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...