3.3 Monitoring of Measurement Process
3 - 52
Lamp Running time
The total Running time (lamp current x usage hours) of
the hollow cathode lamp is calculated. Select an
element to set usage time for. Click [Set] to open
[Running Time] (Fig. 3-40). Enter the number of hours in
the [Running Time] field and click [Set]. If a new lamp is
used, set [Running Time] to 0.
Fig. 3-40 Dialog Box for Setting
the HLamp Running Time
NOTE
: If a new lamp is used, click [Set] with the Running time
being 0.0.
Cuvette Heating Temperature
Indicates the cuvette heating temperature.
The temperature of 0
C means that heating current is
not flowing.
Pressure
Monitor
Represent cooling water flow, argon (Ar) and alternate
gas (Alt) pressure monitors, respectively.
If water or gas is inadequate, the corresponding button
turns red. In the Eco state, the cooling water monitor is
displayed in green. (The alternate gas monitor functions
only when it is set in advance).