(6) Open the [Sample Table] window. Enter 1 for No. of UNK
and assign sample name. Even if multiple samples are
set in the ex-furnace standard addition method, only the
first sample will be measured. Although no sample will be
aspirated actually at the cup position specified for an
unknown sample, input is required for displaying the
sample name.
Fig. 5-8 Sample Table Window
(7) In the [Sample Injection] window in [Autosampler], enter a
value for [Injection Volume]. Usually, a volume of about
20
L is appropriate.