5.5 Twin Injection
5 - 26
5.5 Twin Injection
The twin injection is designed to inject samples through two
injection inlets by using Pyrotube D HR for measurement. It
splits samples in two for injection, enabling you to enlarge the
contact area between the samples and the cuvette surface. You
can reduce the drying time and the incineration time as this
method improves the heat conductive efficiency. You can expect
improved absorption sensitivity in measuring high-melting point
elements such as Mo and V, etc.
(1) Set up a Pyrotube D HR and set the cuvette type on
[Analytical condition] window to “Pyrotube D HR”(Fig. 5-21).
Fig. 5-21 Setting Pyrotube D HR
(2) Set a sample volume to be injected on the [Autosampler]
window. Inject half the volume configured into each port of
the Pyrotube D HR (if 20
μ
L is specified, inject 10
μ
L each
for the total of 20
μ
L).
(3) See “3. Measurement with Conditions Input” for setting
other measurements.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...