3.1 Generation of Measurement Conditions
3 - 4
(1) Click the [Instrument Setup] button (<3> in Fig. 3-3) and the
Instrument Setup window will open.
Fig. 3-3 Instrument Setup Window
(2) The tab for the element selected in the Element window will
be displayed. To change view from one element to another,
click tab (<1> in Fig. 3-3) such as [1-Mg], [2-Cr], [3- Mn] on
the window. The following
(3) Set the following items.
Signal Mode
BKG Correct. is normally used, and need not be
changed.
BKG Correct. : Measurement is made using
background correction signal
(obtained by subtracting reference
signal from sample signal) via the
polarized Zeeman method.
Sample
: Measurement is made using
absorption signal (corresponding to
sample beam in double beam optics)
that has not been corrected.
Reference
: Measurement is made using
background absorption signal
(corresponding to reference beam in
double beam optics).
3.1.4 Instrument
Setup
<1>