4.3.1
4 - 15
1/Conc: W
i
is a reciprocal number of an STD
concentration value for i.
1/Conc2: W
i
is a reciprocal number of a square of an
STD concentration value for i.
Thus, you can create a working curve formula weighted on
low concentration. 1/conc2 has a bigger weight than 1/conc
on low concentration.
(6) Changing the Concentration
The value of concentration can be changed. When
changing it, click the value of [Conc. ] and please input
directly. Click “Recalculate” for the update.
(7) Copying the Working Curve
When clicking the [Copy] button, the working curve graph
can be copied into the clipboard. The graph can be used as
an image file.
(8) Printout of Working Curve
Click the [Print] button, and the working curve graph will be
printed. Usually, it is convenient to print this graph
together with other items by the method described in “4.4.2
Printout of Result.”
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...