3.1 Generation of Measurement Conditions
3 - 6
Fig. 3-4 Wavelength Dialog Box
Moreover, the numerical value can be input directly.
The analytical wavelengths can be checked via [Tools]-
[Analytical Information] on the Method window.
Input range for wavelength is from 190.0 to 900.0 nm.
NOTES
: 1. The Sensitivity Ratio is the rough value (Literature
value) when the highest sensitivity wavelength is
assumed to be "1". There is a difference of
several tens percent depending on the condition,
and use it only as a guide.
2. In same cases, it is necessary to change the slit
width according to wavelength for the separation
of the neighboring line.
Change the slit width referring to the
recommended slit width described in "ANALYSIS
GUIDE (Chapter 7 ELEMENTAL DATA)".
Wavelength
Setting
Normally set at Automatic, and need not be changed.
Automatic:
A peak of lamp energy near the measuring
wavelength is searched for and set automatically.
Direct:
Wavelength is driven directly to the set wavelength.
This mode is dependent on the accuracy of the
wavelength drive mechanism, and will not
necessarily move to the peak position of lamp
energy.
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...