8.6.2
8 - 9
If measurement is stopped immediately though a sample
remains in the graphite cuvette, the next measurement will
be adversely affected. To prevent this, the remaining
sample should be eliminated by performing a dummy
measurement without injecting a sample before resuming
measurement.
NOTE
: Avoid execution of [Maximum Heating] if a sample
remains in the graphite cuvette. Otherwise, the
sample will scatter and contaminate the inside and
vicinity of the cuvette.