6.1 Using Correlation Coefficient of Working Curve (Check Working Curve)
6 - 4
Rerun All STD ..... Remeasures all standard samples
and creates working curve again.
Del. Outer STD ..... Deletes the farthest standard
sample from the working curve
and creates it again.
Run Outer STD .... Remeasures the farthest standard
sample from the working curve
and creates it again.
NOTE
: If the correlation coefficient of the regenerated working
curve is still smaller than the reference value,
measurement will be terminated.
When an element to be measured next is set, however,
it will be measured.
The following messages are printed in the report:
Correlation coefficient is out of range.
Measurement is finished.
Delete a standard sample which deviates most from the
working curve and create a working curve again.
STD
□
is deleted. (
□
represents an STD number deleted.)
Re-measure a standard sample which deviates most from the
working curve and create a working curve again.
Re-measure all the standard samples and create a working
curve again.
6.1.2 Setting
Procedure
Содержание ZA3000 Series
Страница 267: ...3 1 11 3 43 Fig 3 31 Parameter List Window Common conditions The setting of method is thus completed ...
Страница 357: ...5 5 Twin Injection 5 28 Fig 5 24 Dialog Box for Confirming Nozzle Position 7 Click Start measurement ...
Страница 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Страница 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Страница 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Страница 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...