The Pulse subtab can be used to adjust pulse parameters and initiate Pulse mode for
a roller pump or centrifugal pump designated as the primary pump.
1
Roller Pump Pulse Parameters
The Rate, Width, and Base are set up using the Pulse subtab.
To Set the Rate:
1. Touch the
Rate
button.
2. Use the large arrows to change the Rate in increments of 10 beats per minute.
3. Use the small arrows to change the Rate in increments of 1 beat per minute.
4. Touch the
Rate
button again to store the new value.
To Set the Base:
1. Touch the
Base
button.
2. Use the large arrows to change the Base in increments of 10%.
3. Use the small arrows to change the Base in increments of 5%.
4. Touch the
Base
button again to store the new value. The recommended
minimum Base is 30%.
To Set the Width:
1. Touch the
Width
button.
2. Use the large arrows to change the Width in increments of 10%.
3. Use the small arrows to change the Width in increments of 5%.
4. Touch the
Width
button again to store the new value. The recommended
minimum Width is 30%.
See the
Roller Pumps
chapter for details of Pulse mode operation.
1
The use of the term “Pulse mode” refers to the variable pressure and flow
waveforms that result from the cyclic operation of the primary pump. Replication
of a physiologic pulse from a natural heart is not obtained with a pump speed
controller. No claim of clinical benefit from the use of the Pulse mode has been
reviewed by global Regulatory Agencies.
Arterial
16:07:51
Summary
Rate:
Pulse Control:
Pulse
100%
Servo
50%
Base:
Width:
Auto
Auto
Manual
60 BPM
85%
50%
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...