Modules
7-29
Occluder Assembly
Mount the occluder head on a pole according to these installation instructions.
1. Attach the occluder mounting bracket to the desired pole location.
2. Place the ball end of the mounting rod into the occluder head socket.
3. Insert the rod into the mounting bracket and tighten the mounting screws as
needed.
Connect Occluder Module
1. Attach the occluder cable to the occluder module.
2. Insert the occluder module into any base receptacle as close as possible to the
location where the occluder will be used. Arrange cables to minimize exposure
to spills.