Additional Information
12-15
6. Align the studs on the bottom of the tray with the holes in the tray mounting
surface of the base and position the tray against the mounting surface.
7. Install the thumbnuts onto the studs. Make sure the nuts are tight and secure.
Replacing Air Filters and Lamp Bulbs
Air Filters
Base
The fan filter, located on the back panel of the base, should be replaced every 6
months or more often if necessary. Use a flat head screwdriver (provided with
Accessory Kit) to loosen the 2 screws and remove the back panel. Remove the used
filter and insert a new air filter with the blue side of the filter facing base and the white
side facing out. Replace back panel.