Electronic Gas Blender
8-12
System Response to Alarm Conditions
% O2 Icon
Gas Module
Message
Audible Tone
Flashes red
LED flashes red Message in the CCM message area
Dual-tone alarm
System Response to Alert Conditions
Gas Icon
Gas Module
Message
Audible Tone
Flashes yellow LED flashes
Message in the CCM message area
Single-tone alert
yellow
Electronic Gas Blender Messages
Priority Condition
CCM Message
Operator Response
Alarm FiO
2
has dropped below
LOW FiO2
Increase FiO
2
setting to above
minimum limit.
minimum limit to clear alarm.
O
2
source pressure
LOW OXYGEN SUPPLY
Check O
2
hose connections. Check
< 30 psi or more than
PRESSURE O
2
supply and switch source if
18 psi below the air
necessary.
pressure.
O
2
source pressure
HIGH OXYGEN SUPPLY
Check O
2
supply pressure, switch
> 70 psi.
PRESSURE
source if necessary.
Air source pressure
LOW AIR SUPPLY
Check air hose connections. Check
< 30 psi or more than
PRESSURE
air supply and switch source if
18 psi below the O
2
necessary.
pressure.
Air source pressure
HIGH AIR SUPPLY
Check air supply pressure and
> 70 psi.
PRESSURE
switch source if necessary.
95/5 source pressure
LOW 95/5 SUPPLY
Check 95/5 hose connections. Check
< 30 psi or more than
PRESSURE
95/5 supply and switch source if
18 psi below the O
2
necessary.
pressure
95/5 source pressure
HIGH 95/5 SUPPLY
Check 95/5 supply pressure and
> 70 psi
PRESSURE
switch source if necessary.
Alert
Internal malfunction
SERVICE GAS SYSTEM
If not on bypass, calibrate O
2
analyzer.
detected.
If message does not clear or if on
bypass, adjust the FiO
2
and gas flow
using the knobs located on the front of
the system. If the knobs cannot be
turned, use standby gas blender.
Do not use gas system until it has been
serviced by Terumo Customer Service.
See the Messages subtab within the Aux
tab for any additional error or status
messages.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...