Modules
7-32
Occluder Module
Occluder Module Messages
Priority Condition
CCM Message
Operator Response
Alert
Occluder head not
<Occluder name>:
Try disconnecting/reconnecting
moving or not
NOT RESPONDING
occluder head. If message does
responding.
not clear, replace occluder head.
Internal malfunction
<Occluder name>:
Try disconnecting/reconnecting
detected.
SERVICE MODULE
module. If message does not clear,
reassign module.
Error
Occluder becomes
<Occluder name>:
Reconnect and recalibrate occluder
disconnected after
Connect Occluder
head.
calibration.
Calibrated occluder
Calibrate <Occluder Name>
Recalibrate occluder.
loses calibration.
Status Uncalibrated occluder
Calibrate <Occluder Name>
Calibrate occluder.
detected.
Occluder responds to
<Occluder name>: is <open,
Note occluder value, change
an event.
closed, or % flow>
if desired.
Occluder is disconnected <Occluder name>:
Reconnect and calibrate occluder.
Connect Occluder
In the Event of an Occluder Module malfunction:
Symptoms
How to respond
The system responds with indications to the user
that the occluder is non-functional.
•
A ‘connect occluder’ message appears on the
CCM in white text and the occluder maintains
its current position.
•
Audible tone accompanies ‘Service Occluder’
message in flashing yellow text and the
occluder maintains its current position.
Unlatch and open the occluder cover, remove the
tubing, and use a tubing clamp to control flow.
•
Replace the occluder module and reassign a
backup module. Refer to
Reassign Module
Subtab
in the
Central Control Monitor
chapter.
•
If you do not have a backup occluder module,
use increased vigilance throughout the
remainder of the case.
•
If you are using a reservoir-optional circuit:
Any malfunction of a safety system requires
consideration for converting to a circuit with a
venous reservoir.
•
If a venous reservoir is not available, use
extra diligence and be prepared to manually
clamp the line immediately in response to a
pump stop/coast event triggered by an air
alarm.
•
Contact Terumo Customer Service at the end
of the case.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...