Roller Pumps
5-16
Check the Start-up Test
When the system power is first turned on, a self-test is conducted. If the self-test
fails, a message is displayed on the pump display and the CCM. Try turning the
system off then on to repeat the start up check. If the message appears again and
remains on the display, do not use the pump. Contact Terumo Customer Service.
The pump display will be determined by the pump state. Offline indicates the pump
is not assigned to the current Perfusion Screen or a Perfusion Screen has not been
opened. Online indicates the pump is assigned to the current Perfusion Screen.
Offline Display
•
Module ID
•
Pump default direction arrow
•
Last tube size or flow constant entered
•
“READY: CHECK DIRECTION” is displayed when the pump cannot verify that its
default forward direction matches the system direction.
•
No message is displayed if the pump has verified that its default forward direction
matches the system direction.
WARNING:
Always check the direction of rotation before starting a backup or
standby roller pump to make sure that the roller pump default direction matches
the system direction. Retrograde pumping MAY CAUSE SEVERE PATIENT
INJURY OR DEATH by drawing air into the bloodline which might then be
pumped into the patient.
Online Display
•
Pump Name
•
Last tube size or flow constant entered
•
Pump default direction arrow
•
“READY: PRESS START” displayed until user activates controls from pump
control panel or the CCM.
00005
1/2
READY: CHECK DIRECTION
L/min
ARTERIAL
1/2
L/min
READY: PRESS START
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...