Central Control Monitor
4-72
Alarm limits must always be higher than alert limits. The system maintains a
10 mmHg pressure difference between the alert and alarm limit to ensure that the
alert is always less than the alarm.
To change the alarm or alert limit, touch the
Alarm Limit
or
Alert Limit
button and
use the arrows to set a new limit. Touch the
Alarm Limit
or
Alert Limit
button again
to store the new value.
Choose
Next
to continue.
Place the Pressure Sensor
Place the pressure sensor in the location that most closely matches the actual
location of the pressure sensor in the circuit by touching the appropriate white square.
Choose
Do Not Place
to allow placing the sensor at a later time.
Change placement of the sensor by choosing any other white square.
Clamp
Reset
Reset
000:00
Pump
Reset
Reset
000:00
Arterial
Vent
Sucker1
Sucker2
CPG
To Place the Device 'ART Pres',
Touch One of the Valid Location(s) Above:
Do Not Place
Next >
Next >
< Previous
Undo Place
Undo Place
Gas
% O2
FiCO2
0
0
0
Placing a device icon in this
location may interfere
with a CDI display icon.
Set Pressure Limits:
360
Alarm Limit (mmHg)
300
Alert Limit (mmHg)
+ 50 mmHg
+ 10 mmHg
– 10 mmHg
– 50 mmHg
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...