Central Control Monitor
4-41
Priority
Condition
CCM Display Message
Operator Response
Alert
No data from Follower
< Follower Pump Name > : Check the follower pump cable
pump.
NO RESPONSE
connections. Stop the master pump,
open the Master Status subtab within
the master pump tab, and touch
Reconnect button. Start the Master
pump. If Follower pump cannot be
reconnected, do not use the Follower
pump until it has been serviced by
Terumo Cardiovascular Systems.
Reassign pump or use standby
equipment.
No data from Master
< Master Pump Name > : Check the Master Pump cable
pump.
NO RESPONSE
connections. Stop the follower
pump, open the Master Status subtab
within the master pump tab, and touch
the Reconnect button. Start the Master
pump. If Master pump cannot be
reconnected, do not use the pump
until it has been serviced by Terumo
Cardiovascular Systems. Reassign
pump or use standby equipment.
Centrifugal pump has
<Centrifugal Pump Name>: Clamp the lines. Stop and start
detected underspeed
UNDERSPEED
the pump. If the message reappears,
condition.
discontinue use of the pump and
contact Terumo Customer Service.
Reassign pump, use standby
equipment, or manual drive unit.
Pump speed is out of
< Pump Name> :
Check the operation and placement of
servo range limit.
MONITOR FLOW
the servo source. If source is
operating properly, try to initiate
servo operation via the CCM. If
message appears again, discontinue
use of servo operation.
Pump servo range has
< Roller Pump Name> :
Check the operation and placement of
exceeded maximum
MAXIMUM LIMIT
the servo source. If source is
limit.
operating properly, try to initiate
servo operation via the CCM. If
message appears again, discontinue
use of servo operation.
Centrifugal pump servo
<Centrifugal Pump Name>: Check the operation and
range has exceeded
SERVO LIMIT
placement of the servo source.
maximum servo limit.
Alert will clear when speed slows to
within operating range.
Centrifugal pump servo
<Centrifugal Pump Name>: Check operation and placement
range has gone below
SERVO LIMIT
of the servo source. Alert will
minimum servo limit.
clear when speed increases to within
operating
range.
Air sensor malfunction.
< ABD Name > :
Check connection to air sensor, reset
CHECK SENSOR
air sensor, replace if necessary.
Calibrated pressure
< Pressure Name > :
Reconnect and recalibrate pressure
probe disconnected
CONNECT PROBE
probe.
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...