Central Control Monitor
4-13
Pump Controls and Displays
Pump control boxes displayed on the lower third of the CCM can be used to start and
stop pumps and to monitor current pump speed or flow values. Each pump has a
corresponding slide bar located either on the right or left side of the Perfusion Screen
which can be used to adjust the pump speed on the CCM. Pumps can also be
controlled from their front panel controls.
Pump Control Box
All pump control boxes consist of three elements: a Start/Stop button, a speed/flow
display button, and a pump icon button. The only exception is a master/follower pump
control box (see
Roller Pumps
chapter for details of a master/follower pump control
box). The pump type, name, color, display units, and position are all defined in the
CCM Configuration area.
Start Button or Stop Button
The Start button activates and starts the pump at a speed of 0 RPM or a flow of
0.000 L/min. The Stop button stops the pump if it is running and also deactivates it.
If the pump is stopped and deactivated using the Stop button, it must be reactivated
using the Start button for the pump to be used again.
To start a pump:
1. Touch the
Start
button.
•
The button color changes to red and the label changes to Stop.
•
The pump speed/flow display button appears.
•
The pump slide bar appears on one side of the CCM (the primary pump slide bar
is always displayed).
To stop a pump:
1. Touch the
Stop
button.
•
The button color changes to blue and the label changes to Start.
•
If the pump is not the primary pump, the speed/flow button and the pump slide
bar disappear.
Pump Speed/Flow Display
The current value of the pump speed or flow rate is displayed in the units selected
during Configuration.
Touching the button on a primary pump will scroll through three possible display
selections: L/min, L/min/m
2
, or mL/kg/min.
Note:
If a centrifugal pump is not configured to display flow through the use of a
flowmeter, the button will display dashes.
For all other pumps, the pump speed/flow display button appears when the pump is
started and disappears when the pump is stopped. Touching the speed/flow display
button will open the slide bar for that pump and close any open slide bar in the
shared location.
Roller Pump Icon
Arterial
L/min
START
CPG
L/min
STOP
0.000
Centrifugal Pump Icon
Pump Name
Speed/Flow Display
Start/Stop Buttons
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...