Centrifugal Control Unit
6-11
Pulse Mode
The use of the term “Pulse mode” refers to the variable pressure and flow waveforms
that result from the cyclic operation of the primary pump. Replication of a physiologic
pulse from a natural heart is not obtained with a pump speed controller. No claim
of clinical benefit from the use of the Pulse mode has been reviewed by global
Regulatory Agencies
When using Pulse mode during a case, monitor the patient waveform on the
operating room monitor.
A pulse cycle consists of one high speed period (Peak) and one low speed period
(Base). The high speed is controlled by the pump speed control knob or the pump
control slide bar on the CCM and the low speed is determined by the Base setting.
The high speed is displayed on the slide bar and the average speed is displayed in
the pump control box and on the control unit local display. For the centrifugal control
unit, the Width – the amount of time the pulse cycle is at the high speed – is fixed at
50%. The Rate – the number of pulse cycles that occur per minute – is fixed at 60
beats per minute.
Pulse Parameters
The Base must be set up using the CCM:
1. Touch the
Primary Pump
icon to access the Primary Pump tab.
2. Select the
Pulse
subtab.
Arterial
08:22:27
Summary
Rate:
Pulse Control:
Pulse
100%
Servo
50%
Base:
Width:
Auto
Auto
Manual
65%
60 BPM
50%
60 BPM
50%
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...